Carbon isotope analysis of dissolved organic carbon in fresh and saline (NaCl) water via continuous flow cavity ring-down spectroscopy following wet chemical oxidation

January 1, 2015

This work examines the performance and limitations of a wet chemical oxidation carbon analyser interfaced with a cavity ring-down spectrometer (WCO-CRDS) in a continuous flow (CF) configuration for measuring δ13C of dissolved organic carbon (δ13C-DOC) in natural water samples. Low-chloride matrix (22.5 µg C/aliquot) may be analysed. The WCO-CRDS approach requires more total carbon (µg C/aliquot) than conventional CF-isotope ratio mass spectrometer, but is nonetheless applicable to a wide range of DOC concentration and water types, including brackish water, produced water, and basinal brines.

Publication Year 2015
Title Carbon isotope analysis of dissolved organic carbon in fresh and saline (NaCl) water via continuous flow cavity ring-down spectroscopy following wet chemical oxidation
DOI 10.1080/10256016.2015.1009910
Authors Christopher H. Conaway, Randal B. Thomas, Nabil Saad, James J. Thordsen, Yousif K. Kharaka
Publication Type Article
Publication Subtype Journal Article
Series Title Isotopes in Environmental and Health Studies
Index ID 70134310
Record Source USGS Publications Warehouse
USGS Organization National Research Program - Western Branch