Skip to main content
U.S. flag

An official website of the United States government

Improved spectrophotometric cell for hydrothermal solutions

January 1, 1981

A simple, inexpensive spectrophotometric cell was designed for use with aqueous solutions for which temperature is a maximum of 325??C and pressure, 28 MPa. The cell has an internal volume of 5 ml and a path length of 1.31 cm. Each furnace assembly is 120 mm in diameter ?? 150 mm high and will fit into most commercial spectrophotometers. Temperature is controlled by a standard set-point controller and a balancing circuit that is used to maintain the temperature of the sample and reference cell within 1??C of each other at any temperature.

Publication Year 1981
Title Improved spectrophotometric cell for hydrothermal solutions
DOI 10.1063/1.1136597
Authors N.J. Susak, D.A. Crerar, T.C. Forseman, J.L. Haas
Publication Type Article
Publication Subtype Journal Article
Series Title Review of Scientific Instruments
Index ID 70011820
Record Source USGS Publications Warehouse