Screening of ground water samples for volatile organic compounds using a portable gas chromatograph
A portable gas chromatograph was used to screen 32 ground water samples for volatile organic compounds. Seven screened samples were positive; four of the seven samples had volatile organic substances identified by second-column confirmation. Four of the seven positive, screened samples also tested positive in laboratory analyses of duplicate samples. No volatile organic compounds were detected in laboratory analyses of samples that headspace screening indicated to be negative. Samples that contained volatile organic compounds, as identified by laboratory analysis, and that contained a volatile organic compound present in a standard of selected compounds were correctly identified by using the portable gas chromatograph. Comparisons of screened-sample data with laboratory data indicate the ability to detect selected volatile organic compounds at concentrations of about 1 microgram per liter in the headspace of water samples by use of a portable gas chromatograph.
Citation Information
Publication Year | 1989 |
---|---|
Title | Screening of ground water samples for volatile organic compounds using a portable gas chromatograph |
DOI | 10.1111/j.1745-6592.1989.tb01160.x |
Authors | Robert C. Buchmiller |
Publication Type | Article |
Publication Subtype | Journal Article |
Series Title | Ground Water Monitoring Review |
Index ID | 70015118 |
Record Source | USGS Publications Warehouse |