Skip to main content
U.S. flag

An official website of the United States government

Interlaboratory comparison of reference materials for nitrogen-isotope-ratio measurements

January 1, 1995

Aliquots of seven different reference materials were distributed for an interlaboratory comparison of stable nitrogen-isotope-ratio measurements. Results from 15 laboratories were compiled and evaluated selectively to yield provisional values of 515N for each material, i, with respect to atmospheric N2 (o1SN,7air). The 515N values reported by the different laboratories are correlated in such a way that some of the major discrepancies may be removed by normalization (/. e., by altering the length of the ô N scale for each laboratory by an amount defined by local measurements of reference materials with extreme values).

Publication Year 1995
Title Interlaboratory comparison of reference materials for nitrogen-isotope-ratio measurements
Authors John K. Böhlke, Tyler B. Coplen
Publication Type Report
Publication Subtype Other Government Series
Index ID 70188273
Record Source USGS Publications Warehouse